5-14
Table 5-1.Front Panel Controls and Connectors
NO.
NAME
DESCRIPTION
1
LINE switch
This is a two-position switch that can be selected for STBY or OPER. When in
STBY, the power supply provides power to the oven of the high stability time
base to maintain a constant temperature for the crystal. When in OPER, the
power supplies normal power to the unit for operation.
2
RESET
This button, when pressed, resets the display and internal count to zero. When
pressed continuously, this button lights all segments of the LED display and all
annunciator LEDs for a LED test.
FUNCTION
This is a 10-position selector used to select the mode of operation.
FREQ A
This position, when selected, allows the counter to measure frequency at
channel A.
FREQ C
This position, when selected, allows the counter to measure the frequency at
channel C.
PER A
This position, when selected, allows the counter to measure the period at
channel A.
PER AVG A
This position, when selected, allows the counter to make a period average
measurement of the signal at channel A. The number of periods over which the
average is made is determined by the RESOLUTION switch selection.
T. I. AVG A!B
This position, when selected, allows the counter to make a time interval
measurement of a signal applied to channel A. The number of time intervals
over which the average will be made is determined by the RESOLUTION
switch selection made.
T. I. A!B
This position, when selected, allows the counter to make a time interval
measurement. The start signal would be applied to channel A and the stop
signal applied to channel B.
3
CHECK
This position, when selected, applies a 10-MHz signal to the decade counting
assemblies, verifying operation of the SAMPLE RATE control, RESOLUTION
switch, and RESET.
RATIO C/A
This position, when selected, allows the counter to measure the ratio of the
frequency at channel C to the frequency at channel A.
RATIO B/A
This position, when selected, allows the counter to measure the ratio of the
frequency at channel B to the frequency at channel A.
3
Continued
Top
This position is blank and has no function.
This is an eight-position switch that is used to select the resolution in frequency
measurements and N for totalizing and averaging measurements. This also
determines how long the main gate is open for frequency measurements.
4
FREQ
RESOLUTION, N
N
1
10
102
103
104
105
106
107
GATE TIME
(seconds)
1×10-6
10×10-6
100×10-6
1×10-3
.01
.1
1
10
RESOLUTION
(Hz)
1 M
100 k
10 k
1 k
100
10
1
.1
5
SAMPLE RATE
control
This is used to vary the time between measurements continuously from
approximately 2 milliseconds to an indefinite hold of the display.
6
500 MHz, 50 "
This is the channel C input connector.
7, 14
OUTPUT MARKER
These are the channel A and B Schmidt trigger outputs used to indicate when a
channel is triggered by 0 to 300 mV levels with less than 20 nanoseconds delay.